垂直系统
• 输入通道:RTH1002 | 2 oscilloscope channels, 1 multimeter
• RTH:4 oscilloscope channels
• 输入阻抗:1 MΩ ± 1 % || 12 pF ± 2 pF (meas.)
• Analog bandwidth (-3 dB):RTH1002、RTH1004 | ≥ 60MHz
• R&S®RTH1002 with -B221 option and:≥ 100MHz
R&S®RTH1004 with -B241 option
• R&S®RTH1002 with -B222 option and:≥ 200MHz
R&S®RTH1004 with -B242 option
• R&S®RTH1002 with -B223 option and:≥ 350MHz
R&S®RTH1004 with -B243 option
• R&S®RTH1002 with -B224 option and:≥ 500MHz
R&S®RTH1004 with -B244 option
• voltage rating versus frequency,:see figure “Input rating: Maximum signal voltage at
• oscilloscope input”:on page 14.
• Lower frequency limit (-3 dB) at:< 8Hz (meas.)
• AC coupling
• Bandwidth limits:1/2/5/10/20/50/100/200/500kHz, 1/2/5/10/20/50MHz
• only with R&S®RTH-B222, -B242, -B223,:100MHz
-B243, -B224, -B244 options
• only with R&S®RTH-B223, -B243, -B224,:200MHz
-B244 options
• Rise time (calculated):RTH1002、RTH1004 | < 5.8 ns
• R&S®RTH1002 with:B221 option and | < 3.5 ns
• R&S®RTH1004 with:B241 option
• R&S®RTH1002 with:B222 option and | < 1.75 ns
• R&S®RTH1004 with:B242 option
• R&S®RTH1002 with:B223 option and | < 1 ns 2
• R&S®RTH1004 with:B243 option
• R&S®RTH1002 with:B224 option and | < 700 ps 2
• R&S®RTH1004 with:B244 option
• ADC resolution:10 bit
decimation
• input sensitivity > 5 mV/div:±1.5 % of reading
• input sensitivity > 2 mV/div to 5 mV/div:±2 % of reading
• input sensitivity 2 mV/div:±2.5 % of reading
• 输入耦合:DC, AC
• Maximum input voltage:at BNC inputs CAT IV 300 V (RMS), 424 V (peak),
input” on page 14)
• with R&S®RT-ZI10 or R&S®RT-ZI11 probe:CAT IV 600 V, CAT III 1000 V,
derating in line with probe specification
• Position range:±4 div
• Invert:RTH1004 CH1, CH2, CH3, CH4
• RTH:CH1, CH2
• Offset range:input sensitivity ≥ 40 V/div 0 ≥ 1 V/div to ≤ 20 V/div ±200 V ≤ 500 mV/div ±4 V
• ≥ 200MHz (meas.) for input sensitivities ≥ 20 V/div.
• < 1.75 ns (calculated) for input sensitivities ≥ 20 V/div.
• 0.1 div × input sensitivity + 1.5 mV )
input frequency < analog bandwidth
• Common mode rejection (CMRR):DC and AC ≤ 100kHz | > 100 dB (meas.)
水平系统
• Channel deskew:±100 ns
• Reference position:10 %, 50 % or 90 % of measurement
at most 100 000 s
• min.:right edge of measurement display area
• Modes:normal, roll
• Timebase accuracy:±10 ppm
采集系统
• 2 active channels with 2.5 Gsample/s
• 4 active channels with 1.25 Gsample/s
• RTH:1 active channel with 5 Gsample/s 2 active channels with 2.5 Gsample/s
acquisition mode)
• 125 ksample for each channel (high
acquisition mode)
• 62.5 ksample for each channel (high
average acquisition modes)
• 62.5 ksample for each channel (high
average acquisition modes)
• high resolution:average value of samples in decimation
interval
• peak detect:largest and smallest sample in decimation
interval
• envelope:envelope of acquired waveforms;
peak-detect is used.
• average:average of acquired waveforms;
waveforms can be power of 2 from 2 to
• Real-time waveform acquisition rate:max. | 50 000 waveforms/s
(see also R&S®RTH-B1 mixed signal option)
• Trigger level:range | ±4 div from center of screen
• Trigger modes:auto, normal, single
• Trigger sources:RTH1004 CH1, CH2, CH3, CH4
• RTH:CH1, CH2
• Hold-off range:time | 8 ns至10 s, fixed and random
• events:1至1 000 000 000 events
than specified width
• glitch width:800 ps至5000 s
inside or outside the interval
• pulse width:800 ps至5000 s
(SDTV and HDTV require R&S®RTH-K19 option)
• trigger events:all fields, odd fields, even fields, all lines,
line number
• pattern time:800 ps至5000 s
(requires R&S®RTH-K19 option)
• runt pulse width:800 ps至5000 s
• toggle time:800 ps至5000 s
(requires R&S®RTH-K19 option)
• window time:800 ps至5000 s
or either (requires R&S®RTH-K19 option)
• max. data rate:< 250 Mbps
(requires R&S®RTH-K19 option)
• timeout:800 ps至5000 s
(requires R&S®RTH-K19 option)
• interval time:800 ps至5000 s
sources
• RTH:CH1, CH2, CH3, CH4, math
• RTH:CH1, CH2, math
• time based measurements:period, frequency, rise time, fall time,
amplitude, area, overshoot, preshoot,
• AC, DC, AC+DC
• count based measurements:count positive pulses, count negative
edges
• power based measurements:active power, apparent power, reactive
waveforms
• vertical:2 cursors showing time, time difference
and inverse time difference (frequency)
• horizontal:2 cursors showing voltage and voltage
difference
• tracking:vertical cursor additionally showing
waveform
• measure:defines gate for automatic measurements
模板测试
• Sources:RTH1004 CH1, CH2, CH3, CH4, math
• RTH:CH1, CH2, math
• Mask definition:tolerance tube based on analog input
waveform or math waveform
• Number of simultaneous mask tests:up至5
• Actions on violation:none, beep, stop
• History behavior:requires R&S®RTH-K15 option store all
Waveform maths
• Number of math waveforms:1
• Functions:addition, subtraction, multiplication,
square, absolute value, inverse
• Sources:RTH1004 CH1, CH2, CH3, CH4
• RTH:CH1, CH2
FFT
• Sources:RTH1004 CH1, CH2, CH3, CH4
• RTH:CH1, CH2
• FFT sizes:64 ksample or 8 ksample
• Sampling frequency:equal to sampling rate of current
horizontal scale for FFT size of
• 64 ksample;
horizontal scale for FFT size of 8 ksample
• Window types:rectangular, flat-top, Hamming, Hann,
Blackman
• Channel bandwidth:same as bandwidth limits in vertical
additionally accessible over FFT menu
• X-axis scaling:linear or logarithmic
Display characteristics
• Diagram types:Yt, XY, zoom, FFT
• XY mode:parallel display of XY diagram, Xt and Yt
• Interpolation:sin(x)/x
• Persistence:50 ms至10 s; infinite
• Reference signals:up至1 reference signal
Protocol and logic
• Bus trigger and decode:number of bus signals | 1
bus types
• R&S®RTH-K1 option:SPI, I2C
• R&S®RTH-K2 option:UART
• R&S®RTH-K3 option:CAN, LIN
• R&S®RTH-K9 option:CAN-FD
• R&S®RTH-K10 option:SENT
• display types:decoded bus, logical signal, event table
• data format of decoded bus:hex, decimal, binary
results
• Number of simultaneous logging channels:4
• Sources:RTH1004
• oscilloscope mode:up至4 waveform measurements
R&S®RTH1002
• oscilloscope mode:up至4 waveform measurements
• multimeter mode:multimeter measurement
• Measurement speed:1/2/5 measurements/s
• 存储深度:2 Msample per logging channel
• Slot memory:internal memory for up至10 sets of data
Digital voltmeter (DVM)
• Sources:RTH1004 CH1, CH2, CH3, CH4
• Measurements:voltage | DC, AC, AC+DC
with indication of max., min. and average
• Number of active measurements:4
• Maximum resolution:999 counts, 3 digits
Digital multimeter (DMM)
• Sources:RTH1002 multimeter, 4 mm banana inputs, fully
ground
• Measurements:voltage | DC, AC, AC+DC
• 电流:with current clamp or shunt
diode test
• 温度:resistance measurement with PT100 or
capacitance
• Number of active measurements:1
• Maximum resolution:10000 counts, 4 digits
• 输入阻抗:1 V, 10 V | 11.11 MΩ (nom.)
• (voltage DC, AC, AC+DC):100 V 1000 V | 10.10 MΩ (nom.) 10.01 MΩ (nom.)
• Input capacitance:< 100 pF
• Normal mode rejection ratio (NMRR):50Hz/60Hz ± 0.1 % | > 60 dB (meas.)
• Maximum input voltage:CAT III 1000 V (RMS), 1414 V (peak),
stabilization
• DC accuracy:1 V ±(0.05 % + 0.05 % of range) 10 V, 100 V ±(0.05 % + 0.03 % of range) 1000 V ±(0.08 % + 0.03 % of range)
• AC accuracy (AC coupling):1 V, 10 V, 100 V 20 Hz至20kHz ±(0.2 % + 0.05 % of range) 20 kHz至100kHz ±(0.5 % + 0.05 % of range) 1000 V 20 Hz至10kHz ±(0.2 % + 0.05 % of range)
• Resistance ranges:10 % overrange 1.0000 MΩ, 10.000 MΩ, 100.00 MΩ | 1.0000 kΩ, 10.000 kΩ, 100.00 kΩ,
• Resistance accuracy:1 kΩ, 10 kΩ, 100 kΩ, 1 MΩ ±(0.08 % + 0.03 % of range) 10 MΩ ±(0.2 % + 0.05 % of range) 100 MΩ ±(1.5 % + 0.1 % of range)
• Resistance test currents:1 kΩ 10 kΩ 100 kΩ 1 MΩ 10 MΩ 100 MΩ | 1.004 mA (nom.) 101.3 µA (nom.) 10.13 µA (nom.) 1.003 µA (nom.) 100.3 nA (nom.) 100.3 nA || 11.11 MΩ (nom.)
• Continuity range:test current 1.004 mA (nom.), | 1.0000 kΩ
• continuous beep when resistance < 10:Ω
• Continuity accuracy:±(0.1 % + 0.5 Ω)
• Diode test ranges:test current 1.004 mA (nom.) | 3.000 V