电流灵敏度:10 pA
SMU通道数量:4
电压范围:-24 V 至 24 V
最大直流吸收功率:3.6 瓦
PXIe,4通道,±24 V,150 mA,精密PXI源测量单元
ONETEST 欢迎您
厂商登录 | 入驻
登录 | 注册 购物车 微信公众号厂商登录
电流灵敏度:10 pA
SMU通道数量:4
电压范围:-24 V 至 24 V
最大直流吸收功率:3.6 瓦
PXIe,4通道,±24 V,150 mA,精密PXI源测量单元
以下指标取自 Teledyne 官方在线规格书,按原文档小节分组。
| CH <0..3> Output HI | HI force terminal connected to channel power stage (generates and/or dissipates power). Positive polarity is defined as voltage measured on HI > LO. |
| CH <0..3> Guard | Buffered output that follows the voltage of the HI force terminal. Used to drive shield conductors surrounding HI force and Sense HI conductors to minimize effects of leakage and capacitance on low level currents. |
| CH <0..3> Output LO | LO force terminal connected to channel power stage (generates and/or dissipates power). Positive polarity is defined as voltage measured on HI > LO. |
| CH <0..3> Sense HI | Voltage remote sense input terminals. Used to compensate for I*R voltage drops in cable leads, connectors, and switches. |
| NC | No Connect. |
| 10 μA | 10 pA |
| 100 μA | 100 pA |
| 1 mA | 1 nA |
| 10 mA | 10 nA |
| 150 mA | 150 nA |
| 10 μA | ± 100 kΩ |
| 100 μA | ± 10 kΩ |
| 1 mA | ± 1 kΩ |
| 10 mA | ± 100 Ω |
| 150 mA | ± 6.66 Ω |
| Settling time[5] | <100 μs to settle to 0.1% of voltage step, device configured for fast transient response, typical |
| Transient response | <100 μs to recover within ±20 mV after a load current change from 10% to 90% of range, device configured for fast transient response, typical |
| Wideband source noise[6] | 2 mV RMS, typical <20 mVpk-pk, typical |
| Cable guard output impedance | 10 kΩ, typical |
| Remote sense Voltage Add 0.1% of LO lead drop to voltage accuracy specification Current Add 0.03% of range per volt of total HI and LO lead drop to current accuracy specification Maximum lead drop Up to 1 V drop per lead | Voltage |
| Voltage | Add 0.1% of LO lead drop to voltage accuracy specification |
| Current | Add 0.03% of range per volt of total HI and LO lead drop to current accuracy specification |
| Maximum lead drop | Up to 1 V drop per lead |
| Load regulation Voltage 10 μV at connector pins per mA of output load when using local sense, typical Current 20 pA + (10 ppm of range per volt of output change) when using local sense, typical | Voltage |
| Voltage | 10 μV at connector pins per mA of output load when using local sense, typical |
| Current | 20 pA + (10 ppm of range per volt of output change) when using local sense, typical |
| Isolation voltage, channel-to-earth ground[7] | 60 VDC, CAT I, verified by dielectric withstand test, 5 s, continuous |
| Absolute maximum voltage between any terminal and LO | 30 VDC, continuous |
| Available sample rates[8] (600 kS/s)/N where N = 6, 7, 8, … 220 S is samples | Available sample rates[8] |
| Available sample rates[8] | (600 kS/s)/N |
| Sample rate accuracy | ±50 ppm |
| Maximum measure rate to host[9] | 600,000 S/s per channel, continuous |
| Maximum source update rate[10] Sequence length <300 steps per iteration 100,000 updates/s per channel Sequence length ≥300 steps per iteration 100,000 updates/s per board | Sequence length <300 steps per iteration |
| Sequence length <300 steps per iteration | 100,000 updates/s per channel |
| Sequence length ≥300 steps per iteration | 100,000 updates/s per board |
| Input trigger to Source event delay 5 µs Source event jitter 1.7 μs Measure event jitter 1.7 μs | Source event delay |
| Source event delay | 5 µs |
| Source event jitter | 1.7 μs |
| Measure event jitter | 1.7 μs |
| Input triggers Types Start Source Sequence Advance Measure Sources (PXI trigger lines 0 to 7) Polarity Configurable Minimum pulse width 100 ns, nominal Destinations[11] (PXI trigger lines 0 to 7) Polarity Active high (not configurable) Minimum pulse width >200 ns, nominal | Types |
| Types | Start Source Sequence Advance Measure |
| Sources (PXI trigger lines 0 to 7) Polarity Configurable Minimum pulse width 100 ns, nominal | Polarity |
| Polarity | Configurable |
| Minimum pulse width | 100 ns, nominal |
| Destinations[11] (PXI trigger lines 0 to 7) Polarity Active high (not configurable) Minimum pulse width >200 ns, nominal | Polarity |
| Polarity | Active high (not configurable) |
| Minimum pulse width | >200 ns, nominal |
| Output triggers (events) Types Source Complete Sequence Iteration Complete Sequence Engine Done Measure Complete Destinations (PXI trigger lines 0 to 7) Polarity Configurable Pulse width Configurable between 250 ns and 1.6 μs, nominal | Types |
| Types | Source Complete Sequence Iteration Complete Sequence Engine Done Measure Complete |
| Destinations (PXI trigger lines 0 to 7) Polarity Configurable Pulse width Configurable between 250 ns and 1.6 μs, nominal | Polarity |
| Polarity | Configurable |
| Pulse width | Configurable between 250 ns and 1.6 μs, nominal |
| Dimensions | 3U, one-slot, PXI Express/CompactPCI Express module 2.0 cm × 13.0 cm × 21.6 cm (0.8 in. × 5.1 in. × 8.5 in.) |
| Weight | 412 g (14.53 oz) |
| Front panel connectors | 25-position D-SUB, male |
| Operating | 0 °C to 55 °C |
| Storage | -40 °C to 70 °C |
| Operating | 10% to 70%, noncondensing. Derate 1.3% per °C above 40 °C. |
| Storage | 5% to 95%, noncondensing. |
| Pollution degree | 2 |
| Maximum altitude | 2,000 m (800 mbar) (at 25 °C ambient temperature) |
| Operating vibration | 5 Hz to 500 Hz, 0.3 g RMS |
| Non-operating vibration | 5 Hz to 500 Hz, 2.4 g RMS |
| Operating shock | 30 g, half-sine, 11 ms pulse |
官方规格书:https://www.ni.com/docs/en-US/bundle/pxie-4143-specs/page/specs.html
NI官方渠道
官方渠道(官网 / 热线 / 邮箱)
官方渠道(中国大陆)
• 官网 www.ni.com/zh-cn
• 在线选购 www.ni.com/zh-cn/shop.html
• 经销商查询
官网「查找经销商」 www.ni.com/zh-cn/shop/advisors.html
• 联系方式
NI 中文官网不公布中国区电话与邮箱,报价、技术支持、校准、维修一律通过官网在线表单提交(「联系我们」页)
全国办事处
公司总部
• NI(National Instruments)
• 地址
11500 N Mopac Expwy, Austin, TX 78759-3504, United States
• 母公司
Emerson(2023 年完成收购,NI 隶属艾默生测试与测量业务集团)