SMU通道数量:12
电压范围:-24 V 至 24 V
最大直流吸收功率:1.2 瓦
最大直流供电功率:2.4 瓦
PXIe,12通道,±24 V,精密PXI源测量单元
ONETEST 欢迎您
厂商登录 | 入驻
登录 | 注册 购物车 微信公众号厂商登录
SMU通道数量:12
电压范围:-24 V 至 24 V
最大直流吸收功率:1.2 瓦
最大直流供电功率:2.4 瓦
PXIe,12通道,±24 V,精密PXI源测量单元
以下指标取自 Teledyne 官方在线规格书,按原文档小节分组。
| 376734J-01 | October 2026 |
| 376734H-01 | June 2026 |
| 376734G-01 | April 2025 |
| 376734F-01 | March 2025 |
| 376734E-01 | August 2024 |
| 376734D-01 | June 2023 |
| 376734C-01 | November 2021 |
| 376734B-01 | December 2018 |
| 376734A-01 | January 2018 |
| PXIe-4162 (10 pA) | NI Measurement & Automation Explorer (MAX) |
| Device Front Panel | PXIe-4162 12-CH 10pA SMU |
| PXIe-4162 (100 pA) | NI Measurement & Automation Explorer (MAX) |
| Device Front Panel | PXIe-4162 12-CH Precision SMU |
| CH <0..11> Sense LO | Voltage remote sense input terminals. Used to compensate for I*R voltage drops in cable leads, connectors, and switches. |
| CH <0..11> Guard | Buffered output that follows the voltage of the HI force terminal. Used to drive shield conductors surrounding HI force and Sense HI conductors to minimize effects of leakage and capacitance on low level currents. |
| CH <0..11> Sense HI | Voltage remote sense input terminals. Used to compensate for I*R voltage drops in cable leads, connectors, and switches. |
| CH <0..11> Output HI | HI force terminal connected to channel power stage (generates and/or dissipates power). Positive polarity is defined as voltage measured on HI > LO. |
| CH <0..11> Output LO | LO force terminal connected to channel power stage (generates and/or dissipates power). Positive polarity is defined as voltage measured on HI > LO. |
| Calibration HI | For external calibration use only, otherwise leave unconnected. |
| NC | No Connect. |
| 0 through 11 | ±24 V |
| ≥58 W | 2.4 W |
| <58 W | 1.4 W |
| 1 μA | 10 pA |
| 10 μA | 100 pA |
| 100 μA | 1 nA |
| 1 mA | 10 nA |
| 10 mA | 100 nA |
| 60 mA or 100 mA[10] | 1 μA |
| 10 μA | 100 pA |
| 100 μA | 1 nA |
| 1 mA | 10 nA |
| 10 mA | 100 nA |
| 60 mA or 100 mA[15] | 1 µA |
| 1 μA [16] | 0 to ±5 MΩ |
| 10 μA | 0 to ±500 kΩ |
| 100 μA | 0 to ±50 kΩ |
| 1 mA | 0 to ±5 kΩ |
| 10 mA | 0 to ±500 Ω |
| 60 mA [17] | 0 to ±80 Ω |
| 100 mA [18] | 0 to ±50 Ω |
| Settling time[23] | <500 μs, typical[24] |
| Transient response[25] | <100 μs, typical[26] |
| Voltage | No additional error due to lead drop |
| Current | No additional error due to lead drop |
| Maximum lead drop | 1 V drop/lead |
| Voltage[27] | 50 μV/mA, typical |
| Current | (30 pA + 20 ppm of range)/volt, typical |
| Overcurrent or overvoltage | Automatic shutdown, output disconnect relay opens |
| Overtemperature | Automatic shutdown, output disconnect relay opens |
| From Sense HI, Sense LO, or Guard when VOutput HI > 0 V | -0.5 V to (VOutput HI + 0.5 V) |
| From Sense HI, Sense LO, or Guard when VOutput HI ≤ 0 V | (VOutput HI - 0.5 V) to 0.5 V |
| From all other pins | ±25 V |
| Available sample rates[30] | (600 kS/s)/N, nominal where N = 6, 7, 8, … 220 S is samples |
| Sample rate accuracy | ±50 ppm |
| Maximum measure rate to host[31] | 100,000 S/s per channel, continuous, nominal |
| Single channel | 100,000 updates/s, nominal |
| All channels simultaneously | 40,000 updates/s per channel, nominal |
| Source event delay | 8.5 µs, nominal |
| Source event jitter | 1.7 μs, nominal |
| Measure event jitter | 1.7 μs, nominal |
| Types | Start Source Sequence Advance Measure |
| Polarity | Active high (not configurable) |
| Minimum pulse width | 100 ns |
| Polarity | Active high (not configurable) |
| Minimum pulse width | >200 ns |
| Types | Source Complete Sequence Iteration Complete Sequence Engine Done Measure Complete |
| Polarity | Active high (not configurable) |
| Pulse width | 230 ns |
| +3.3 V | Idle |
| +3.3 V | Full Output Load |
| +12 V | Idle |
| +12 V | Full Output Load |
| +3.3 V | Idle |
| +3.3 V | Full Output Load |
| +12 V | Idle |
| +12 V | Full Output Load |
| Dimensions | 3U, one-slot, PXI Express/CompactPCI Express module 2.1 cm × 13.1 cm × 21.4 cm (0.8 in. × 5.1 in. × 8.4 in.), nominal For more information, visit ni.com/dimensions and search by module number. |
| Weight | 394 g (13.9 oz), nominal |
| Front panel connector | Custom 62-position D-SUB, female |
| Operating temperature for chassis with slot cooling capacity ≥58 W[32] | 0 °C to 55 °C |
| Operating temperature for all other compatible chassis | 0 °C to 40 °C |
| Storage | -40 °C to 71 °C |
| Operating | 10% to 90%, noncondensing |
| Storage | 5% to 95%, noncondensing |
| Pollution Degree | 2 |
| Maximum altitude | 2,000 m (800 mbar) (at 25 °C ambient temperature) |
| Operating vibration | 5 Hz to 500 Hz, 0.3 g RMS |
| Non-operating vibration | 5 Hz to 500 Hz, 2.4 g RMS |
| Operating shock | 30 g, half-sine, 11 ms pulse |
官方规格书:https://www.ni.com/docs/en-US/bundle/pxie-4162-specs/page/specs.html
NI官方渠道
官方渠道(官网 / 热线 / 邮箱)
官方渠道(中国大陆)
• 官网 www.ni.com/zh-cn
• 在线选购 www.ni.com/zh-cn/shop.html
• 经销商查询
官网「查找经销商」 www.ni.com/zh-cn/shop/advisors.html
• 联系方式
NI 中文官网不公布中国区电话与邮箱,报价、技术支持、校准、维修一律通过官网在线表单提交(「联系我们」页)
全国办事处
公司总部
• NI(National Instruments)
• 地址
11500 N Mopac Expwy, Austin, TX 78759-3504, United States
• 母公司
Emerson(2023 年完成收购,NI 隶属艾默生测试与测量业务集团)
0条提问
X
提问
X
回答
发表口碑