扫描速率:140 周期/秒
模式(n线):1线
总线连接器:PXI混合
最大直流切换电流:2 A
32通道,300 V,2线PXI多路复用器开关模块
ONETEST 欢迎您
厂商登录 | 入驻
登录 | 注册 购物车 微信公众号厂商登录
扫描速率:140 周期/秒
模式(n线):1线
总线连接器:PXI混合
最大直流切换电流:2 A
32通道,300 V,2线PXI多路复用器开关模块
以下指标取自 Teledyne 官方在线规格书,按原文档小节分组。
| CHx | Channel terminals |
| COM0+ | Routing destination for all channel terminals |
| COM0- | Routing destination for the 1WREF0 lead |
| CH0 through CH31 | Channel terminals that route to COM0+ |
| CH 32 through CH63 | Channel terminals that route to COM1+ |
| COM0+ | Routing destination for CH0 through CH31 |
| COM1+ | Routing destination for CH32 through CH63 |
| COM0- | Routing destination for the 1WREF0 lead |
| COM1- | Routing destination for the 1WREF1 lead |
| CH0+ through CH31+ | Channel terminals that route to COM0+ |
| CH0- through CH31- | Channel terminals that route to COM0- |
| COM0+ | Routing destination for all positive channels |
| COM0- | Routing destination for all negative channels |
| CH0+ through CH15+ | Channel terminals that route to COM0+ |
| CH0- through CH15- | Channel terminals that route to COM0- |
| CH16+ through CH31+ | Channel terminals that route to COM1+ |
| CH16- through CH31- | Channel terminals that route to COM1- |
| COM0+ | Routing destination for CH0+ through CH15+ |
| COM0- | Routing destination for CH0- through CH15- |
| COM1+ | Routing destination for CH16+ through CH31+ |
| COM1- | Routing destination for CH16- through CH31- |
| CH0A+ through CH15A+ | Channel terminals that route to COM0+ |
| CH0A- through CH15A- | Channel terminals that route to COM0- |
| CH0B+ through CH15B+ | Channel terminals that route to COM1+ |
| CH0B- through CH15B- | Channel terminals that route to COM1- |
| COM0+ | Routing destination for all positive channels (A and B) |
| COM0- | Routing destination for all negative channels (A and B) |
| CHx | Channel terminals |
| COM0+ | Routing destination for all channel terminals |
| COM0- | Routing destination for the 1WREF0 lead |
| Maximum switching voltage[1] Channel-to-channel 300 V Channel-to-ground 300 V, CAT I | Channel-to-channel |
| Channel-to-channel | 300 V |
| Channel-to-ground | 300 V, CAT I |
| Maximum switching power (per channel) AC systems 60 W, 62.5 VA (up to 60 Hz) DC systems Refer to the following figure. | AC systems |
| AC systems | 60 W, 62.5 VA (up to 60 Hz) |
| DC systems | Refer to the following figure. |
| Maximum total current (switching or carry) | 2 A |
| Minimum switch load[2] | 20 mV /1 mA |
| DC path resistance[3] Initial <1 Ω, warranted End-of-life ≥2 Ω | Initial |
| Initial | <1 Ω, warranted |
| End-of-life | ≥2 Ω |
| Differential thermal EMF | 2.5 μV, typical[4] <12 μV, maximum |
| Channel-to-channel DC leakage at 300 V | 500 GΩ |
| Bandwidth (-3 dB, 50 Ω termination) 1-wire >30 MHz 2-wire >25 MHz | 1-wire |
| 1-wire | >30 MHz |
| 2-wire | >25 MHz |
| Channel-to-channel isolation (50 Ω termination) (1-wire and 2-wire) 10 kHz >80 dB 100 kHz >60 dB 1 MHz >40 dB | 10 kHz |
| 10 kHz | >80 dB |
| 100 kHz | >60 dB |
| 1 MHz | >40 dB |
| Open channel isolation (50 Ω termination) (1-wire and 2-wire) 10 kHz >80 dB 100 kHz >60 dB 1 MHz >40 dB | 10 kHz |
| 10 kHz | >80 dB |
| 100 kHz | >60 dB |
| 1 MHz | >40 dB |
| Mechanical | 1 × 108 cycles |
| Electrical | 300 VDC, 60 mADC resistive |
| 30 VDC, 2 ADC resistive | 1 × 105 cycles |
| <30 mV, <10 mA | 2.5 × 106 cycles |
| Sources | PXI trigger lines <0...7> |
| Minimum pulse width | 150 ns |
| Destinations | PXI trigger lines <0...7> |
| Default pulse width | 2 µs, typical |
| Relay type | Electromechanical, latching |
| Relay contact material | Palladium-ruthenium, gold covered |
| I/O connector | 100-position HDI right angle, male |
| Power requirement PXI 6 W at 5 V 2.5 W at 3.3 V PXI Express 7.5 W at 12 V 2.5 W at 3.3 V | PXI |
| PXI | 6 W at 5 V 2.5 W at 3.3 V |
| PXI Express | 7.5 W at 12 V 2.5 W at 3.3 V |
| Dimensions (L × W × H) | 3U, one slot, PXI/cPCI module 21.6 cm × 2.0 cm × 13.0 cm (8.5 in. × 0.8 in. × 5.1 in.) |
| Weight | 209 g (7.4 oz) |
| Operating temperature | 0 °C to 55 °C |
| Storage temperature | -20 °C to 70 °C |
| Relative humidity | 5% to 85%, noncondensing |
| Pollution Degree | 2 |
| Maximum altitude | 2,000 m |
| Operational Shock | 30 g peak, half-sine, 11 ms pulse (Tested in accordance with IEC 60068-2-27. Test profile developed in accordance with MIL-PRF-28800F.) |
| Random vibration, operating | 5 Hz to 500 Hz, 0.3 grms |
| Random vibration, nonoperating | 5 Hz to 500 Hz, 2.4 grms (Tested in accordance with IEC 60068-2-64. Nonoperating test profile exceeds the requirements of MIL-PRF-28800F, Class 3.) |
官方规格书:https://www.ni.com/docs/en-US/bundle/pxi-2527-specs/page/specs.html
NI官方渠道
官方渠道(官网 / 热线 / 邮箱)
官方渠道(中国大陆)
• 官网 www.ni.com/zh-cn
• 在线选购 www.ni.com/zh-cn/shop.html
• 经销商查询
官网「查找经销商」 www.ni.com/zh-cn/shop/advisors.html
• 联系方式
NI 中文官网不公布中国区电话与邮箱,报价、技术支持、校准、维修一律通过官网在线表单提交(「联系我们」页)
全国办事处
公司总部
• NI(National Instruments)
• 地址
11500 N Mopac Expwy, Austin, TX 78759-3504, United States
• 母公司
Emerson(2023 年完成收购,NI 隶属艾默生测试与测量业务集团)
0条提问
X
提问
X
回答
发表口碑
AI问答