RF功率计最大频率:6 GHz
RF功率计最小频率:50 MHz
动态范围:-40 dBm 至 23 dBm
测量速度:15 S/s
6 GHz,-40 dBm ~ +23 dBm功率范围,RF功率传感器设备
ONETEST 欢迎您
厂商登录 | 入驻
登录 | 注册 购物车 微信公众号厂商登录
RF功率计最大频率:6 GHz
RF功率计最小频率:50 MHz
动态范围:-40 dBm 至 23 dBm
测量速度:15 S/s
6 GHz,-40 dBm ~ +23 dBm功率范围,RF功率传感器设备
以下指标取自 Teledyne 官方在线规格书,按原文档小节分组。
| Green | The USB-5680 is connected to a computer or a chassis and is functioning properly. |
| Yellow | An error has occurred. |
| Off | The USB-5680 is not initialized. |
| Frequency range | 50 MHz to 6 GHz |
| Dynamic range | -40 dBm to +23 dBm |
| Input return loss 50 MHz to <2 GHz <-26 dB 2 GHz to 6 GHz <-20 dB | 50 MHz to <2 GHz |
| 50 MHz to <2 GHz | <-26 dB |
| 2 GHz to 6 GHz | <-20 dB |
| Measurement range Measurement range 1 -40 dBm to <-5 dBm Measurement range 2 -5 dBm to +23 dBm | Measurement range 1 |
| Measurement range 1 | -40 dBm to <-5 dBm |
| Measurement range 2 | -5 dBm to +23 dBm |
| >10 | >10.0 |
| 5 | 3.2 |
| 0 | 1.0 |
| -5 | 0.32 |
| -10 | 0.10 |
| -15 | 0.032 |
| -20 | 0.010 |
| -25 | 0.0032 |
| -30 | 0.0010 |
| -35 | 0.00032 |
| -40 | 0.00010 |
| >10 | >10.0 |
| 5 | 3.2 |
| 0 | 1.0 |
| -5 | 0.32 |
| -10 | 0.10 |
| -15 | 0.032 |
| -20 | 0.010 |
| -25 | 0.0032 |
| -30 | 0.0010 |
| -35 | 0.00032 |
| -40 | 0.00010 |
| Signal-channel bandwidth | 100 Hz |
| Linearity Power level <18 dBm ±0.13 dB Power level ≥18 dBm ±0.18 dB | Power level <18 dBm |
| Power level <18 dBm | ±0.13 dB |
| Power level ≥18 dBm | ±0.18 dB |
| Calibration factor[1] | ±0.06 dB |
| -40 to <-5 dB | <2.5 nW |
| -5 to +23 dBm | <0.6 μW |
| Zero set[3] -40 to <-5 dBm <10 nW -5 to +23 dBm <1.7 μW | -40 to <-5 dBm |
| -40 to <-5 dBm | <10 nW |
| -5 to +23 dBm | <1.7 μW |
| Zero drift[4] -40 to <-5 dBm .<3.0 nW -5 to +23 dBm < 0.5 μW | -40 to <-5 dBm |
| -40 to <-5 dBm | .<3.0 nW |
| -5 to +23 dBm | < 0.5 μW |
| Temperature compensation[5] 0 °C to 50 °C ±0.06 dB 20 °C to 30 °C 0 dB | 0 °C to 50 °C |
| 0 °C to 50 °C | ±0.06 dB |
| 20 °C to 30 °C | 0 dB |
| Effect of digital modulation Power level <18 dBm ±0.02 dB Power level ≥18 dBm ±0.10 dB | Power level <18 dBm |
| Power level <18 dBm | ±0.02 dB |
| Power level ≥18 dBm | ±0.10 dB |
| Measurement | True root mean square/Average power |
| Measurement resolution | 0.01 dB |
| Offset range | .±100 dB |
| Averaging range[6] | 1 to 256 |
| Measurement speed Low-aperture time mode 15 measurements per second High-aperture time mode 1 measurement per second | Low-aperture time mode |
| Low-aperture time mode | 15 measurements per second |
| High-aperture time mode | 1 measurement per second |
| Interface | USB 2.0 and 1.1 compliant |
| Maximum DC voltage at RF port | 25 V |
| Absolute power | 33 dBm |
| Typical current | 100 mA |
| Maximum current | 150 mA |
| Dimensions | 8.5 cm × 3.0 cm × 5.6 cm (3.35 in × 1.18 in × 2.2 in) |
| Weight | 180 g (6.4 oz) |
| Maximum altitude | 2,000 m (800 mbar) (at 25 °C ambient temperature) |
| Pollution Degree | 2 |
| Ambient temperature range | 0 to 55 °C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2.) |
| Relative humidity range[7] (noncondensing) At 55 °C 45% At 40 °C 75% At 30 °C 95% | At 55 °C |
| At 55 °C | 45% |
| At 40 °C | 75% |
| At 30 °C | 95% |
| Ambient temperature range | -51 °C to +71 °C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2.) |
| Relative humidity range | 5% to 95%, noncondensing (Tested in accordance with IEC 60068-2-56.) |
| Operating shock | 30 g peak, half-sine, 11 ms pulse |
| Random vibration Operating 5 Hz to 500 Hz, 0.3 grms Nonoperating 5 Hz to 500 Hz, 2.4 grms | Operating |
| Operating | 5 Hz to 500 Hz, 0.3 grms |
| Nonoperating | 5 Hz to 500 Hz, 2.4 grms |
官方规格书:https://www.ni.com/docs/en-US/bundle/usb-5680-specs/page/specs.html
NI官方渠道
官方渠道(官网 / 热线 / 邮箱)
官方渠道(中国大陆)
• 官网 www.ni.com/zh-cn
• 在线选购 www.ni.com/zh-cn/shop.html
• 经销商查询
官网「查找经销商」 www.ni.com/zh-cn/shop/advisors.html
• 联系方式
NI 中文官网不公布中国区电话与邮箱,报价、技术支持、校准、维修一律通过官网在线表单提交(「联系我们」页)
全国办事处
公司总部
• NI(National Instruments)
• 地址
11500 N Mopac Expwy, Austin, TX 78759-3504, United States
• 母公司
Emerson(2023 年完成收购,NI 隶属艾默生测试与测量业务集团)
0条提问
X
提问
X
回答
发表口碑
AI问答