福禄克校准|Fluke Calibration
5606 Full Immersion PRT 详细指标
可在 -200 °C 至 160 °C 的温度范围内运行。 · 准确度 ±0.05 °C。 · 这些 PRT 的过渡连接处和引线均能够承受全量程的 PRT 温度。
详细指标
以下指标取自福禄克校准(FLUKE CALIBRATION)官网该产品页「技术指标」表,为官网公布值;实际精度与适用条件以官方手册为准。
General Specifications
• Temperature range:-200°C to 160°C
• Nominal resistance at 0.01°C:100 Ω ±0.1 Ω
• Temperature coefficient:0.00385 Ω/Ω/°C
• Accuracy 1:±0.04°C at 0.01°C ±0.06°C at 160°C
• Short-term repeatability 2:±0.03°C at 0.01°C ±0.04°C at 160°C
• Drift 3:±0.03°C at 0.01°C ±0.04°C at 160°C
• Hysteresis:±0.015°C
• Sheath length:50 mm ±5 mm (2 in ±0.2 in)
• Sheath diameter:3.1 mm ±0.1 mm (1/8 in ±0.004 in)
• Sheath material:316 SST
• Transition junction temperature range 4:-200°C to 160°C
• Transition junction dimensions:No transition
• Sensor length:30 mm ±3mm (1.2 in ±0.1 in)
• Sensor location:3 mm ±1 mm from tip (1.2 in ±0.04 in)
• Minimum insulation resistance:20 MΩ at 23°C
• Minimum immersion length:Full immersion
• Maximum immersion depth in liquid medium:Full immersion
• Maximum immersion depth in dry Medium:Full immersion
• Response time 5:12 seconds typical
• Self heating (in 0°C bath):±0.003°C
• Lead-wire cable type:Enameled copper wire
• Lead-wire length:2.4 meters (8 ft)
• Lead-wire temperature range:160°C
• Calibration:Calibration not included. NVLAP accredited calibration available. See ordering information.
Calibration Uncertainty
• Temperature:-197°C 1924 Calibration Uncertainty 6 0.03°C Tolerance w/out Calibration 7 0.4°C
型号与选件
• 5606-50-P:
• 5606 PRT 5606 Full Immersion PRT:
• 1930-4-10 系统对比校准,-200 至 157°:
• 5606-50-X:
• 1930-5-10 系统对比校准,-40 至 157:
资源下载
以上规格取自福禄克校准(FLUKE CALIBRATION)官网 5606 Full Immersion PRT 产品页技术指标表。
福禄克校准|Fluke Calibration · onetest 仪器资源库