是德科技 | KEYSIGHT
E4991B 1MHz至3GHz 分析仪 详细指标
High-performance benchtop impedance analyzers
– 1 MHz to 3 GHz (Option
300)
– 1 MHz to 1 GHz (Option
100)
– 1 MHz to 500 MHz (Option
050)
– without Option E4991B-1E5:
–
± 20 ppm (5 °C to 40 °C)
– with Option E4991B-1E5:
– with Option E4991B-1E5:
–
± 0.5 ppm/year (5 °C to 40 °C) (typical)
– Power (when
50 Ω load is connected to test port):
– –40 dBm to 1 dBm
– 电流 (when 短路 connected 至测试端口)
– 0.0894 mArms to 10 mArms
– 电压 (when 开路 connected 至测试端口)
– 4.47 mVrms to 502 mVrms
Accuracy: (Power, when
50 Ω load is connected to test port)
– Minimum: –3 dB, Maximum: +2 dB (23°C
± 5°C)
– Minimum: –5 dB, Maximum
+4 dB (5 °C to 40 °C)
– Minimum: –4 dB, Maximum: +3 dB (23°C
± 5°C)
– Minimum: –6 dB, Maximum
+5 dB (5 °C to 40 °C)
条件:Output impedance: 50 Ω (nominal)
DC Bias (Option E4991B-001)
Output impedance (series)
15 Ω (typical)
条件:1. When the unit is set at mV or mA, the entered value is rounded to 0.1 dB resolution.
– ± {0.05% + 5 mV + (|Idc[mA]| x 20 Ω)} (23 °C
± 5 °C)
– ± {0.2% + 10 mV + (|Idc[mA]| x
40 Ω)} (5 °C to 40 °C)
Current limit range
1 mA to 100 mA (both source and sink are limited to same current.)
Current limit resolution
2 µA
Current limit accuracy
± 4% (5 °C to 40 °C, typical)
Output impedance (shunt)
20 kΩ minimum (typical)
– ± {0.2% + 20 µA + (|Vdc[V]|/10 kΩ)} (23 °C
± 5 °C)
– ± {0.4% +
40 µA + (|Vdc[V]|/5 kΩ)} (5 °C to 40 °C)
Voltage limit range
0.3 V to 40 V (both positive and negative sides are limited to same voltage.)
Voltage limit resolution
1 mV
Voltage limit accuracy: ± (2% + 20 mV + |Idc| x
20 Ω) (5 °C to 40 °C, typical)
– ± {0.2% + 10 mV + (|Idc[mA]| x
2 Ω)}
– ± {0.8% + 24 mV + (|Idc[mA]| x
4 Ω)}
(5 °C to 40 °C, typical)
– ± {0.2% +
25 µA + (|Vdc[V]|/40 k Ω)}
– ± {0.8% +
60 µA + (|Vdc[V]|/20 k Ω)}
(5 °C to 40 °C, typical)
Linear frequency, log frequency, OSC level (voltage, current, power), DC bias (voltage, current),
Number of measurement points
2 to 1601
– 类型: point 延迟, 扫描延迟, segment 延迟
Available setup parameters for each segment
条件:(power, voltage, or current), DC bias (voltage or current), segment time, segment delay.
Sweep frequency range, number of measurement points, point averaging factor, oscillator level
Number of segments
1 to 201
条件:Conditions for defining accuracy
Temperature: 23 °C ± 5 °C
1
Accuracy-specified plane
7-mm connector of test head
– Same points at which the calibration is done.
2
Basic accuracy (Typical)
0.45%
Accuracy when open/short/load calibration is performed 1. If the calibration is performed in 5 °C to 18 °C or 28 °C to 40 °C, the accuracy is degraded to doubled value (typical). 2. If the calibration is performed in different frequency points or different DC bias points from the measurement
精度 degraded 至 doubled 值 (典型值)
|Z|, |Y|: ±(Ea + Eb) [%] (see Figures 1 through
4 for examples of calculated accuracy)
L, C, X, B: ± (Ea + Eb) x √(1 + D2x) [%]
R, G: ± (Ea + Eb) x √(1 + Q2x) [%]
±
especially at 10
x
(1 + D2
Ea + Eb (1 + Q2 x)tan
Accuracy when open/short/load/low-loss capacitor calibration is
–23 dBm ≤ oscillator level ≤
+1 dBm
Calibration points are same as measurement points (User frequency mode)
Measurement is performed within
± 1 °C from the calibration temperature
条件:calibration is performed at 23 °C ± 5 °C. When the calibration is performed beyond 23 °C
Ea = (Within ± 5 °C from the calibration temperature. Measurement accuracy applies when the
± 5 °C, measurement error doubles.)
At –23 dBm ≤ oscillator level ≤
1 dBm:
– 0.60 [%] (1 MHz ≤ Frequency ≤
100 MHz)
– 0.70 [%] (100 MHz < Frequency ≤
500 MHz)
– 1.00 [%] (500 MHz < Frequency ≤
1 GHz)
– 2.00 [%] (1 GHz < Frequency ≤
1.8 GHz)
– 4.00 [%] (1.8 GHz < Frequency ≤
3 GHz)
At –33 dBm ≤ oscillator level <
–23 dBm:
– 0.65 [%] (1 MHz ≤ Frequency ≤
100 MHz)
– 0.75 [%] (100 MHz < Frequency ≤
500 MHz)
– 1.05 [%] (500 MHz < Frequency ≤
1 GHz)
– 2.05 [%] (1 GHz < Frequency ≤
1.8 GHz)
– 4.05 [%] (1.8 GHz < Frequency ≤
3 GHz)
L, C, X, B: ± √(Ea + Eb)2 + (Ec Dx)2 [%]
R, G: ± √(Ea + Eb)2 + (EcQx)2 [%]
±
especially at 10
±Q2
x
(1 + Q2
Ec (1 + D2 x)tan
+ 0.03 [%] at |Zx| < 1 Ω
1000 |Zx|
+ 0.03 [%] at |Zx| < 3 Ω
1000 |Zx|
+ 0.03 [%] at |Zx| < 1 Ω
1000 |Zx|
+
[%] at |Zx | > 1.8 kΩ 1000
60000
+
[%] at |Zx | > 600 Ω 1000
60000
+
[%] at |Zx | > 1.8 kΩ 1000
60000
[%] at 1 Ω ≤ |Zx| ≤ 1.8 kΩ
1000
[%] at 3 Ω ≤ |Zx| ≤ 600 Ω
1000
[%] at 1 Ω ≤ |Zx| ≤ 1.8 kΩ
1000
At –40 dBm ≤ oscillator level <
–33 dBm:
– 0.80 [%] (1 MHz ≤ Frequency ≤
100 MHz)
– 0.90 [%] (100 MHz < Frequency ≤
500 MHz)
– 1.20 [%] (500 MHz < Frequency ≤
1 GHz)
– 2.20 [%] (1 GHz < Frequency ≤
1.8 GHz)
– 4.20 [%] (1.8 GHz < Frequency ≤
3 GHz)
At 1 MHz ≤ frequency ≤
10 MHz
At 10 MHz < frequency <
100 MHz
At 100 MHz ≤ frequency ≤
3 GHz
Zs = (Specification values of “point averaging factor ≥
8” is applied only when point averaging
factors at both calibration and measurement are
8 or greater.)
At oscillator level =
–3 dBm or –13 dBm:
– (11 + 0.5 × F) [mΩ] (averaging factor ≥
8)
– (12 + 0.5 × F) [mΩ] (averaging factor ≤
7)
At oscillator level =
–23 dBm:
– (12 + 0.5 × F) [mΩ] (averaging factor ≥
8)
– (16 + 0.5 × F) [mΩ] (averaging factor ≤
7)
At –23 dBm < oscillator level ≤
1 dBm:
– (17 + 0.5 × F) [mΩ] (averaging factor ≥
8)
– (21 + 0.5 × F) [mΩ] (averaging factor ≤
7)
At –33 dBm ≤ oscillator level <
–23 dBm:
– (25 + 0.5 × F) [mΩ] (averaging factor ≥
8)
– (50 + 0.5 × F) [mΩ] (averaging factor ≤
7)
At –40 dBm ≤ oscillator level <
–33 dBm:
– (50 + 0.5 × F) [mΩ] (averaging factor ≥
8)
– (100 + 0.5 × F) [mΩ] (averaging factor ≤
7)
Yo = (Specification values of “point averaging factor ≥
8” is applied only when point averaging
factors at both calibration and measurement are
8 or greater.)
At –17 dBm ≤ oscillator level ≤
1 dBm:
– (1.7 + 0.1 × F) [μS] (averaging factor ≥
8)
– (4.0 + 0.1 × F) [μS] (averaging factor ≤
7)
At –23 dBm ≤ oscillator level <
–17 dBm:
– (4.0 + 0.1 × F) [μS] (averaging factor ≥
8)
– (8.0 + 0.1 × F) [μS] (averaging factor ≤
7)
At –33 dBm ≤ oscillator level <
–23 dBm:
– (10.0 + 0.1 × F) [μS] (averaging factor ≥
8)
– (30.0 + 0.1 × F) [μS] (averaging factor ≤
7)
At –40 dBm ≤ oscillator level <
–33 dBm:
– (20.0 + 0.1 × F) [μS] (averaging factor ≥
8)
– (60.0 + 0.1 × F) [μS] (averaging factor ≤
7)
Figure
1. |Z|, |Y| Measurement accuracy when open/short/load calibration
is performed. Oscillator level = –13 dBm
–3 dBm. Point averaging factor
≥ 8 within
± 5 °C from the calibration temperature.
Figure
2. |Z|, |Y| Measurement accuracy when open/short/load
calibration is performed. Oscillator level –13 dBm
–3 dBm. Point
averaging factor ≤ 7 within
± 5 °C from the calibration temperature.
Calculated 阻抗测量精度 (Continued)
Figure
5. Q accuracy without low-loss capacitor calibration
(Specification) and with low-loss capacitor calibration (Typical).
条件:kind of calibration data. The reference plane is called the calibration reference plane.
– Connect open, short, and load standards to the desired reference plane and measure each
Low-loss capacitor calibration
– Connect the dedicated standard (low-loss capacitor) to the calibration reference plane and
Port extension compensation (fixture selection)
条件:plane, set the electrical length between the calibration plane and the device contact.
– When a device is connected to a terminal that is extended from the calibration reference
Select the model number of the registered test fixtures in the E4991B’s setup toolbar or
enter the electrical length for the user’s test fixture
Figure
3. |Z|, |Y| Measurement accuracy when open/short/load calibration
is performed. Oscillator level = –33 dBm. Point averaging factor ≥
8 within
± 5 °C from the calibration temperature.
Figure
4. |Z|, |Y| Measurement accuracy when open/short/load
calibration is performed. Oscillator level =
–33 dBm. Point averaging
factor ≤ 7 within
± 5 °C from the calibration temperature.
– When a device is connected to a terminal that is extended from the calibration reference
plane, make open and/or short states at the device contact and measure each kind of
– Obtain calibration/compensation data at fixed frequency covering the entire frequency
range of the E4991B. In device measurement, calibration or compensation is applied
to each measurement point by using interpolation. Even if the measurement points
are changed by altering the sweep setups, you don’t need to retake the calibration/
条件:data become invalid and retaking calibration/compensation data is recommended.
– Obtain calibration/compensation data at the same frequency as used in actual device
measurement, which are determined by the sweep setups. Each set of calibration/
compensation data is applied to each measurement at the same frequency point. If the
measurement points are changed by altering the sweep setups, calibration/compensation
– Internal, external (external trigger input connector), bus (GPIB/LAN/USB), manual (front
– Sweep-to-sweep averaging
1 to 999 (integer)
– Point averaging
1 to 999 (integer)
– Type/size
10.4 inch TFT color LCD
– Resolution: XGA (1024 x 768)
1
– Data trace
4 data traces per channel (maximum)
– Memory trace
4 memory traces per channel (maximum)
– Data + Memory, Data - Memory, Data x Memory, Data/ Memory, Offset, Equation Editor
– For scalar parameters: linear Y-axis, log Y-axis
条件:1. Valid pixels are 99.99% and more. Below 0.01% of fixed points of black, green, or red are not regarded as
– For complex parameters: Z, Y, εr, µr: polar, complex; Γ: polar, complex, Smith, admittance
– Each measurement channel has a display window with independent stimulus. Up to 4
– 10 independent markers per trace. Reference marker available for delta marker operation
– Search type: max value, min value, multi-peak, multi-target, peak, peak left, peak right,
target, target left, target right, and width parameters with userdefined bandwidth values
– Marker continuous mode, Δ marker mode, Marker coupled mode, Marker value substitution
(Marker&), Marker zooming, Marker list, Marker statistics, and Marker signal/dc bias
– 3-component model (4 models)
– 4-component model (3 models)
– Equivalent circuit parameters calculation, frequency characteristics simulation
Define the test limit lines that appear on the display for define the test limit lines that appear
on the display for pass/fail testing. Defined limits may be any combination of horizontal/sloping
lines and discrete data points. testing. Defined limits may be any combination of horizontal/
GPIB
24-pin D-Sub (Type D-24), female; compatible with IEEE-488.
IEEE-488 interface specification is designed to be used in environment where electrical noise is
relatively low. LAN or USBTMC interface is recommended to use at the higher electrical noise
Universal serial bus jack, Type A configuration; female; provides connection to mouse,
Universal serial bus jack, Type B configuration (4 contacts inline); female; provides connection
to an external PC; compatible with USBTMC-USB488 and USB
2.0.LA USB Test and
Measurement Class (TMC) interface that communicates over USB, complying with the IEEE
488.1 and IEEE
488.2 standards.
Handler interface
36-pin centronics, female
Connector type
7-mm connector
条件:Input impedance: 50 Ω (nominal)
Frequency: 10 MHz
± 10 ppm (typical)
Level: 0 dBm
± 3 dB (typical)
条件:Output impedance: 50 Ω (nominal)
Frequency: 10 MHz
± 10 ppm (typical)
Level: 0 dBm ± 3 dB into
50 Ω (typical)
条件:Output impedance: 50 Ω (nominal)
High stability frequency reference output connector (Option E4991B-1E5)
– LOW threshold voltage
0.5 V
– HIGH threshold voltage
2.1 V
– Input level range
0 V to +5 V
条件:Non-operating storage condition
– 20% to 80% at wet bulb temperature <
+29 °C (non-condensation))
Altitude
0 m to 2,000 m (0 feet to 6,561 feet)
Vibration: 0.21 Grms maximum
5 Hz to 500 Hz
Temperature
–10 °C to +60 °C
条件:Figure 6. Definition of pulse width (Tp).
– 20% to 90% at wet bulb temperature <
+40 °C (non-condensation)
Altitude
0 m to 4,572 m (0 feet to 15,000 feet)
Vibration: 2.1 Grms maximum
5 Hz to 500 Hz
条件:frequency is identical to the transmitted interference signal test frequency.
frequency is identical to the transmitted interference signal test frequency.
European Council Directive
2004/108/EC
– 4 kV CD / 8 kV AD
IEC
61000-4-3:2006 +A1:2007 +A2:2010
EN
61000-4-3:2006 +A1:2008 +A2:2010
– 3 V/m, 80-1000 MHz, 1.4 to 2.0 GHz / 1V/m, 2.0 to 2.7 GHz
80% AM
IEC
61000-4-4:2004 +A1:2010
EN
61000-4-4:2004 +A1:2010
– 1 kV power lines / 0.5 kV signal lines
– 0.5 kV line-line / 1 kV line-ground
NOTE-1: When tested at 3 V/m according to EN61000-4-3, the measurement accuracy will
be within specifications over the full immunity test frequency range except when the analyzer
NOTE-2: When tested at 3 V according to EN61000-4-6, the measurement accuracy will be
within specifications over the full immunity test frequency range except when the analyzer
ICES-001:2006 Group
1, Class A
AS/NZS CISPR11:2004
适用型号 KN11, KN61000-6-1 和 KN61000-6-2
European Council Directive
2006/95/EC
IEC 61010-1:2010 / EN
61010-1:2010
EMC, safety, environment and compliance (Continued)
CAN/CSA C22.2 No.
61010-1-12
This product complies with the WEEE Directive (2002/96/EC) marking requirements. The
affixed label indicates that you must not discard this electrical/electronic product in domestic
To return unwanted products, contact your local Keysight ofice .
Product Category: With reference to the equipment types in the WEEE Directive Annex I, this
product is classed as a “Monitoring and Control instrumentation” product. Do not dispose in
Power requirements 90 V to 264 V AC (Vpeak > 120 V), 47 Hz to 63 Hz
300 VA maximum
Main unit: See Figure 7 through Figure
9
Option 007 test head dimensions: See Figure
11
Option 010 test head dimensions: See Figure
12
Figure
7. Main unit dimensions (front view, in millimeters).
Marker/Analysis Instrument State
M/µ654
x 1+/-.0Meas Format
Impedance Analyzer E4991B
1 MHz-3 GHz
374159
Figure
8. Main unit dimensions (rear view, in millimeters).
WNANAL YZERF36
ICES/NMB-001
100 - 240 VLINE
RSVD 145 1923391927 22130 9
12
Figure
9. Main unit dimensions (side view, in millimeters).
Only for E4991B
E4991B T est Head
±42 V Peak Max
Figure
10. Test head dimensions (in millimeters).
Figure
11. Option E4991B-007 test head dimensions (in millimeters).
Temperature Characteristic Test KitTemperature Characteristic Test Kit
Avoid Static DischargeAvoid Static Discharge ±42V
±42V
Figure
12. Option E4991B-010 test head dimensions (in millimeters).
Only for E4991B
E4991B-010 T est Head
±42 V Peak Max
Option E4991B-002 Material Measurement (Typical)
Permeability parameters: |µ r|, µr’, µr”, tanδ
Using with Keysight 16453A
1 MHz to 1 GHz (typical)
Using with Keysight 16454A
1 MHz to 1 GHz (typical)
– Open, short, and load calibration at electrodes of 16453A when using the
16453A
– Open, short, and load calibration at test port (7-mm connector) of the test head, then Short
compensation at 16454A when using the
16454A
条件:– Calibration is performed at an environmental temperature within the range of 23 °C ± 5 °C.
– Measurement accuracy doubles when calibration temperature is
5 °C to 18 °C or 28 °C to
40 °C.
条件:is performed at 23 °C ± 5 °C. When the calibration is below 18 °C or above 28 °C,
– Temperature deviation: within
± 5 °C from the calibration temperature
– Environment temperature: Measurement accuracy applies when the calibration
条件:Ea = at Frequency ≤ 1 GHz: 1. In fixed frequency calibration mode, if a measurement frequency point is not included in the calibration points, the accuracy at the measurement point is degraded to its doubled value (typical).
– Same as calibration points
1
– Point averaging factor: ≥
8
– Electrode pressure setting of 16453A: maximum
Typical accuracy of permittivity parameters: εr’ accuracy (at tanδ <
0.1)
Loss tangent accuracy of ε ‘r (= ∆tanδ): ± (Ea + Eb) (at tanδ <
0.1) where
=
:
+ 0.25
+
[%] f ['r m t 1– 13
2 f √['r m
0.002 + 0.001
Option E4991B-002 Material Measurement (Typical) (Continued)
Eb = f = Measurement frequency [GHz] t = Thickness of MUT (material under test) [mm]
[‘r m = Measured value of [‘r tand = Measured value of dielectric loss tangent
Typical accuracy of permeability parameters: µr’ accuracy (at tand <
0.1)
Loss tangent accuracy of µ r (= ∆tand): ±( Ea + Eb ) (at tand <
0.1) where
F = h = Height of MUT (material under test) [mm] b = Inner diameter of MUT (material under test) [mm] c = Outer diameter of MUT (material under test) [mm]
µ’r m = Measured value of µ’r tand = Measured value of loss tangent
• 1 + ['r m
0.002 tand
=
µ' r m 4 + 0.02 25 + Fµ'r m 1 + 15 2 f
2 [%] f Fm´r m Fµ'r m
×
条件:Figure 13. Permittivity accuracy (∆[‘r) vs. frequency (at t = 0.3 mm, typical).
∆µr m' • tand µ'r m
100 h ln c [mm] b
Option E4991B-002 Material Measurement (Typical) (Continued)
条件:1. This graph shows only frequency dependence of Ea to simplify it. The typical accuracy of tan d is defined as Ea + Eb; refer to “Typical accuracy of permittivity
Option E4991B-002 Material Measurement (Typical) (Continued)
Figure
16. Dielectric loss tangent (tan d) accuracy vs. frequency
(at t = 0.3 mm, typical)
1
Figure
17. Dielectric loss tangent (tan d) accuracy vs. frequency
Figure
18. Dielectric loss tangent (tan d) accuracy vs. frequency
Figure 19. Permittivity ([' r) vs. frequency (at t =
0.3 mm, typical)
Figure 20. Permittivity ([’ r) vs. frequency (at t =
1 mm, typical)
Figure 21. Permittivity ([’ r) vs. frequency (at t =
3 mm, typical)
Figure 22. Permeability accuracy (∆µ'r) vs. frequency (at F =
0.5 mm, typical)
Figure 24. Permeability accuracy (∆µ'r) vs. frequency (at F =
10 mm, typical)
Option E4991B-002 Material Measurement (Typical) (Continued)
Examples of calculated permeability measurement accuracy
Figure 23. Permeability accuracy (∆µ'r) vs. frequency (at F =
3 mm, typical)
条件:1. This graph shows only frequency dependence of Ea to simplify it. The typical accuracy of tan δ is defined as Ea + Eb; refer to “Typical accuracy of permeability
Figure
25. Permeability loss tangent (tan d) accuracy vs. frequency
(at F = 0.5 mm, typical)
1
Figure
26. Permeability loss tangent (tan d) accuracy vs. frequency
Figure
27. Permeability loss tangent (tan d) accuracy vs. frequency
Figure 28. Permeability (µ' r) vs. frequency (at F =
0.5 mm, typical)
Figure 29. Permeability (µ' r) vs. frequency (at F =
3 mm, typical)
Figure 30. Permeability (µ' r) vs. frequency (at F =
10 mm, typical)
Option E4991B-002 Material Measurement (Typical) (Continued)
Option E4991B-007 Temperature Characteristic Test Kit
This section contains specifications and supplemental information for the E4991B Option
E4991B-007. Except for the contents in this section, the E4991B standard specifications
– –55 °C to +150 °C (at the test port of the high temperature cable)
– +5 °C to +40 °C (Main unit, test head, and their connection cable)
– 1 MHz to 3 GHz (Option
300)
– 1 MHz to 1 GHz (Option
100)
– 1 MHz to 500 MHz (Option
050)
条件:Measurement accuracy (at 23 °C ± 5 °C)
– The measurement accuracy is specified when the following conditions are met:
Source power accuracy at the test port of the high temperature cable:
– Minimum: –4 dB, Maximum: +2 dB (23°C
± 5°C)
– Minimum: –6 dB, Maximum
+4 dB (5 °C to 40 °C)
– Minimum: –5 dB, Maximum: +3 dB (23°C
± 5°C)
– Minimum: –7 dB, Maximum
+5 dB (5 °C to 40 °C)
– Calibration: open, short and load calibration is completed at the test port (7-mm
– Calibration temperature: calibration is performed at an environmental temperature
within the range of 23 °C
± 5 °C. Measurement accuracy doubles when calibration
temperature is
+5 °C to +18 °C or +28 °C to +40 °C.
– Measurement temperature range: Within
± 5 °C of calibration temperature
阻抗, admittance 和相位 angle 精度
条件:1. The high temperature cable must be kept at the same position throughout calibration and
|Z|, |Y| ± (Ea + Eb) [%] (see Figure 31 through Figure
34 for calculated accuracy) q ± (Ea + Eb)
Ea = At –23 dBm ≤ oscillator level ≤
1 dBm:
0.70 [%] (1 MHz ≤ ƒ≤
100 MHz)
0.80 [%] (100 MHz < IJ
500 MHz)
1.10 [%] (500 MHz < IJ
1 GHz)
2.10 [%] (1 GHz < IJ
1.8 GHz)
4.10 [%] (1.8 GHz < IJ
3 GHz)
At –33 dBm ≤ oscillator level <
–23 dBm:
0.75 [%] (1 MHz ≤ ƒ≤
100 MHz)
0.85 [%] (100 MHz < IJ
500 MHz)
1.15 [%] (500 MHz < IJ
1 GHz)
2.15 [%] (1 GHz < IJ
1.8 GHz)
4.15 [%] (1.8 GHz < IJ
3 GHz)
At –40 dBm ≤ oscillator level <
–33 dBm:
0.90 [%] (1 MHz ≤ ƒ≤
100 MHz)
1.00 [%] (100 MHz < IJ
500 MHz)
1.30 [%] (500 MHz < IJ
1 GHz)
2.30 [%] (1 GHz < IJ
1.8 GHz)
4.30 [%] (1.8 GHz < IJ
3 GHz)
Eb = Zs + Yo × |Zx| ×
100 [%]
Zs = At oscillator level =
–3 dBm, or –13 dBm:
(23 + 0.5 × F) [mΩ] (point averaging factor ≥
8)
(24 + 0.5 × F) [mΩ] (point averaging factor ≤
7)
At oscillator level =
–23 dBm:
(24 + 0.5 × F) [mΩ] (point averaging factor ≥
8)
(28 + 0.5 × F) [mΩ] (point averaging factor ≤
7)
At –23 dBm < oscillator level ≤
1 dBm:
(29 + 0.5 × F) [mΩ] (point averaging factor ≥
8)
(36 + 0.5 × F) [mΩ] (point averaging factor ≤
7)
At –33 dBm ≤ oscillator level <
–23 dBm:
(35 + 0.5 × F) [mΩ] (point averaging factor ≥
8)
(70 + 0.5 × F) [mΩ] (point averaging factor ≤
7)
At –40 dBm ≤ oscillator level <
–33dBm:
(50 + 0.5 × F) [mΩ] (point averaging factor ≥
8)
(150 + 0.5 × F) [mΩ] (point averaging factor ≤
7)
(Where, F is frequency in MHz)
Yo = At –17 dBm ≤ oscillator level ≤
1 dBm:
(8 + 0.1 × F) [μS] (averaging factor ≥
8)
(10 + 0.1 × F) [μS] (averaging factor ≤
7)
Option E4991B-007 Temperature Characteristic Test Kit (Continued)
Option E4991B-007 Temperature Characteristic Test Kit (Continued)
At –23 dBm ≤ oscillator level <
–17 dBm:
(10 + 0.1 × F) [μS] (averaging factor ≥
8)
(14 + 0.1 × F) [μS] (averaging factor ≤
7)
At –33 dBm ≤ oscillator level <
–23 dBm:
(15 + 0.1 × F) [μS] (averaging factor ≥
8)
(40 + 0.1 × F) [μS] (averaging factor ≤
7)
At –40 dBm ≤ oscillator level <
–33 dBm:
(35 + 0.1 × F) [μS] (averaging factor ≥
8)
(80 + 0.1 × F) [μS] (averaging factor ≤
7)
Calculated 阻抗/Admittance 测量精度
Figure
34. |Z|, |Y| measurement accuracy when open/short/load
calibration is performed. Oscillator level =
–33 dBm. Point averaging
factor ≤ 7 within
± 5 °C of calibration temperature.
Figure
32. |Z|, |Y| measurement accuracy when open/short/load
calibration is performed. Oscillator level –13 dBm
–3 dBm. Point
averaging factor ≤ 7 within
± 5 °C of calibration temperature.
Figure
31. |Z|, |Y| measurement accuracy when open/short/load
calibration is performed. Oscillator level = –13 dBm
–3 dBm. Point
averaging factor ≥ 8 within
± 5 °C of calibration temperature.
Figure
33. |Z|, |Y| measurement accuracy when open/short/load
calibration is performed. Oscillator level =
–33 dBm. Point averaging
factor ≥ 8 within
± 5 °C of calibration temperature.
When the temperature at the test port (7-mm connector) of the high temperature cable
changes from the calibration temperature, typical measurement accuracy involving
temperature dependence effects (errors) is applied. The typical measurement accuracy
is represented by the sum of error due to temperature coefficients ( E´a, Y´o and Z´s),
hysteresis error ( Eah , Yoh and Zsh) and the specified accuracy.
– Temperature compensation data is acquired at the same temperature points as
条件:Typical measurement accuracy (involving temperature dependence effects):
|Z|, |Y|: ± ( Ea + Eb + Ec + Ed) [%] q : ± (Ea + Eb + Ec + Ed)
Where, Ea, Eb = Refer pages
25 and 26.
Ed = Z´s× ∆T + Zsh + (Y´ o× ∆T + Yoh) × |Zx| ×
100 [%]
Here, E´a, Z´s and Y´o are given by the following equations
Without temperature compensation With temperature compensation
1 MHz ≤ ƒ < 500 MHz 500 MHz ≤ ƒ ≤
3 GHz
Ea´ 0.006 + 0.015 × ƒ [%/°C] 0.006 + 0.015 × ƒ [%/°C] 0.006 +
0.015 × ƒ [%/°C]
Zs´ 1 + 10 × ƒ [mΩ/°C] 1 + 10 × ƒ [mΩ/°C] 5 +
2 × ƒ [mΩ/°C]
Yo´ 0.3 + 3 × ƒ [µS/°C] 0.3 + 3 × ƒ [µS/°C] 1.5 +
0.6 × ƒ [µS/°C]
Figure
35. Typical frequency characteristics of temperature coefficient
(Ec+Ed)/∆T, when |Zx|=
10 Ω and 250 Ω.
Typical Effects of Temperature Change on Measurement Accuracy
Eah, Zsh and Yoh are given by following equations
条件:∆T = Difference of measurement temperature-from calibration temperature.
when the E4991B with Option 002 and
Eah = Ea´ × ∆Tmax ×
0.3 [%]
Zsh = Zs´ × ∆Tmax ×
0.3 [mΩ]
Yoh = Yo´ × ∆Tmax ×
0.3 [µS]
Use ∆T = 0 °C if temperature compensation is set to off and the difference
≤ 5 °C.
Use ∆T = 0 °C if temperature compensation is set to on and the difference
≤ 20 °C.
∆Tmax = Maximum temperature change (°C) at the test port from calibration
temperature after the calibration is performed. Use ∆Tmax =
0 °C if
maximum temperature change ≤
10 °C.
Typical Effects of Temperature Change on Measurement Accuracy (Continued)
Typical Material Measurement Accuracy When Using Options
002 and 007
Material measurement accuracy contains the permittivity and permeability measurement accuracy
以上指标摘自厂家官方数据手册,完整定义、测试条件与保证值口径以手册为准。完整指标、测试条件与定义口径请见产品页「软件和手册 | 白皮书」栏目中的官方数据手册。